Auto ELISA Processor BIOBASE1001
Features:
1. Adopt the principle of pressure type liquid detection.
2. With TIP detection, clot detection function.
3. High-precision pipetting, the lowest volume can reach 10μl.
4. Position sensing, real-time monitoring of gripper status.
5. Multi-spacing and multi-strength adjustable grippers can be adapted to grasp and place microplates and plate covers of different sizes.
6. Dual thread control, can run simultaneously with the sampling arm without interfering with each other.
Technical Parameters:
Model
BIOBASE1001
Sample&Reagent Unit
Sample Racks
6*20 push-pull sample rack
Sampling Channel
1 channel
Liquid Level Detection
Air pressure level detection, clot detection function
Tip
1000μl, 2 tip racks
Pipetting
10μl~1000μl
Pipetting Precision
100μl: CV≤1%, Accuracy≤2.5%
1000μl: CV≤0.5%, Accuracy≤1%
Reagent Position
2 reagent racks
Dilution Plate Position
1 dilution plate position with incubation and shaking
Washing Unit
Washer
1 unit
Washing Head
16-channel(32-pin), double row cleaning, hanging installation, with leakage recovery
Injection Uniformity
CV<1.5%
Washing Residual
<2μl
Reading Unit
Reading Channels
8 independent photometric channels, mono and dichromatic reading
Absorbance Range
0.000~4.000Abs
Linear Range
0.000~3.000Abs
Repeatability
CV≤1.0%
Stability
≤±0.003Abs
Spectral Range
400~700nm
Optical Filters
4 standard filters(405, 450, 492, 630nm)
Total 10 filters positions
Reading Accuracy
≤±0.02Abs from 0.0 to 1.0 OD
≤±0.03Abs from 1.0 to 2.0 OD
Sensitivity
≥0.01mg/L
Channel Difference
≤0.02Abs
Incubation Unit
Incubators
2 shaking incubators
Cover Position
1(2 covers)
Heating Method
Dry heating with cover
Shaking Function
Horizontal circular oscillation, 1~3 levels adjustable, amplitude 2.5mm
Temperature Range
RT to 60 °C
Temperature Uniformity
±0.5℃
Temperature Fluctuation
±0.5℃
Data Management
Software
Above Windows 7 system
LIS System
Bi-direction, support HL7 protocol
Results
Absorbance and results reviewable by software
Printer
External optional
Built-in Scanner
Optional
Other Function
Biosafety Protection
Optional(Negative pressure working zone, HEPA filter)
Disinfection
UV lamp
Lighting
LED
IAP Function
Firmware can be upgraded online at any time
Working Condition
Power Supply
AC220V±10%, 50/60Hz(Standard); 110V±10%, 60Hz(Optional)
Temperature
10~40℃
Humidity
30~80%
Size&Weight
External Size(W*D*H)
870*680*870; 465*475*700mm
Net Weight
155kg; 55kg
Package Size(W*D*H)
1050*850*1150mm
1100*820*750mm
Gross Weight
210kg; 105kg
Model
BIOBASE1001
Sample&Reagent Unit
Sample Racks
6*20 push-pull sample rack
Sampling Channel
1 channel
Liquid Level Detection
Air pressure level detection, clot detection function
Tip
1000μl, 2 tip racks
Pipetting
10μl~1000μl
Pipetting Precision
100μl: CV≤1%, Accuracy≤2.5%
1000μl: CV≤0.5%, Accuracy≤1%
Reagent Position
2 reagent racks
Dilution Plate Position
1 dilution plate position with incubation and shaking
Washing Unit
Washer
1 unit
Washing Head
16-channel(32-pin), double row cleaning, hanging installation, with leakage recovery
Injection Uniformity
CV<1.5%
Washing Residual
<2μl
Reading Unit
Reading Channels
8 independent photometric channels, mono and dichromatic reading
Absorbance Range
0.000~4.000Abs
Linear Range
0.000~3.000Abs
Repeatability
CV≤1.0%
Stability
≤±0.003Abs
Spectral Range
400~700nm
Optical Filters
4 standard filters(405, 450, 492, 630nm)
Total 10 filters positions
Reading Accuracy
≤±0.02Abs from 0.0 to 1.0 OD
≤±0.03Abs from 1.0 to 2.0 OD
Sensitivity
≥0.01mg/L
Channel Difference
≤0.02Abs
Incubation Unit
Incubators
2 shaking incubators
Cover Position
1(2 covers)
Heating Method
Dry heating with cover
Shaking Function
Horizontal circular oscillation, 1~3 levels adjustable, amplitude 2.5mm
Temperature Range
RT to 60 °C
Temperature Uniformity
±0.5℃
Temperature Fluctuation
±0.5℃
Data Management
Software
Above Windows 7 system
LIS System
Bi-direction, support HL7 protocol
Results
Absorbance and results reviewable by software
Printer
External optional
Built-in Scanner
Optional
Other Function
Biosafety Protection
Optional(Negative pressure working zone, HEPA filter)
Disinfection
UV lamp
Lighting
LED
IAP Function
Firmware can be upgraded online at any time
Working Condition
Power Supply
AC220V±10%, 50/60Hz(Standard); 110V±10%, 60Hz(Optional)
Temperature
10~40℃
Humidity
30~80%
Size&Weight
External Size(W*D*H)
870*680*870; 465*475*700mm
Net Weight
155kg; 55kg
Package Size(W*D*H)
1050*850*1150mm
1100*820*750mm
Gross Weight
210kg; 105kg
Auto ELISA Processor BIOBASE1001 — professional solution in the ELISA Equipment category
Auto ELISA Processor BIOBASE1001 is a modern business-class equipment designed for daily operation in demanding conditions. The combination of reliability, accuracy and durability makes this product the optimal choice for companies seeking consistent process quality, predictable results and reduced operating costs. The purchase price is a reasonable investment in the development and scaling of the enterprise.
Why Auto ELISA Processor BIOBASE1001
- Reliability and durability: industrial-grade design and materials withstand intense loads.
- Accuracy and repeatability: stable performance that reduces the risk of defects and downtime.
- Process efficiency: optimizing staff time and increasing productivity at every stage.
- Compliance with standards: manufactured taking into account current international requirements for the business segment.
- Ease of integration: Auto ELISA Processor BIOBASE1001 fits easily into existing process lines and IT infrastructure.
Areas of application ELISA Equipment
ELISA Equipment solutions are widely used in manufacturing, research laboratories, construction sites, energy and agricultural sectors, pharmaceutical and food industries. Implementing Auto ELISA Processor BIOBASE1001 allows you to increase productivity, ensure accuracy of operations and optimize resource utilization across the entire enterprise.
Key benefits for business
Return on investment
Reduced service costs and downtime, predictive budget planning, fast ROI thanks to stable operation.
Scalability
Readiness to expand capacity and integrate with other solutions without re-engineering processes.
Safety and quality control
Compliance with standards, which simplifies audits and increases the trust of customers and partners.
Technical highlights
- Industrial reliability of components and assemblies for long-term trouble-free operation.
- Stable parameters that ensure repeatability of results and product quality.
- Energy efficiency and optimized life cycle cost of ownership.
- Compatibility with related equipment and standard data exchange protocols.
Service and support
Together with Auto ELISA Processor BIOBASE1001 you receive professional advice on selection, commissioning and personnel training. We provide warranty and post-warranty service, prompt delivery of consumables and spare parts. Our team supports projects from the procurement stage to the stable operation of the complex so that your business receives the expected result every day.
To select a package for your needs or to receive a commercial offer, add Auto ELISA Processor BIOBASE1001 to the cart or contact our manager. We will help integrate ELISA Equipment into existing processes without downtime and unnecessary costs.
No questions about this product, be the first and ask your question.