AC Series Class II B2 Biological Safety Cabinet BSC-1100IIB2-X BSC-1300IIB2-X BSC-1500IIB2-X BSC-1800IIB2-X
Features:
1. Color key display: large screen, more clear and dynamic display of various values of the safety cabinet, and has the function of appointment timing.
2. Automatically adjust the wind speed.
3. With power-failure memory function, the state before power-failure will be restored after power-on, and there will be visual prompts and alarms.
4. The integrated design of the operation area makes it easy to disassemble and clean.
5. The cabinet is separated from the base. The height of the adjustable base can be customized.
6. Interlocking function: UV lamp and front window; UV lamp and fan, LED lamp; fan and front window.
7. It has the function of appointment timing, which can automatically set the time of starting and shutting down the safety cabinet and the running time of ultraviolet lamps, fans and sockets.
8. The front window closes the double trigger signal, so that the sterilization and disinfection function of the ultraviolet lamp is normally turned on.
9. The negative pressure air duct is equipped with a foreign matter filtering structure to prevent foreign matter such as paper scraps from entering the fan system and affecting the normal operation of the product.
10. The front window glass has a one-key lift function.
11. The cabinet is designed with a 10° inclination angle, which is in line with ergonomic principles, with a larger viewing angle, convenient operation and more humanization.
12. The one-piece hand rest has a large fat surface, which has a larger contact area with the arm and is more comfortable.

Technical Parameters:
Model
BSC-1100IIB2-X
BSC-1300IIB2-X
BSC-1500IIB2-X
BSC-1800IIB2-X
External Size(W*D*H)
1100*775*2250mm
1300*775*2250mm
1500*775*2250mm
1800*775*2250mm
Internal Size(W*D*H)
940*600*660mm
1150*600*660mm
1350*600*660mm
1625*600*660mm
Tested Opening
Safety height 200mm(8")
Max Opening
480mm
Inflow Velocity
0.53±0.025m/s
Downflow Velocity
0.33±0.025m/s
ULPA Filter
Two, 99.9995% efficiency at 0.12μm, filter life indicator
Front Window
Motorized. Two-layer laminated toughened glass≥6mm. Anti UV
Noise
≤65dB
UV Lamp
30W*1
30W*1
40W*1
40W*1
UV timer, UV life indicator, emission of 235.7 nanometers for most efficient decontamination
LED Lamp
12W*2
14W*2
16W*2
16W*2
Illumination
≥1000Lux
Consumption
1300W
1300W
1300W
1300W
Waterproof Socket
Two, total load of two sockets: 500W
Display
LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and downflow velocity, filter life, humidity and temperature, system working time etc.
Control System
Microprocessor
Airflow System
0% air recirculation, 100% air exhaust
Visual and Audio Alarm
Abnormal airflow velocity, Filter replacement, Front window at unsafe height, High filter pressure alarm, abnormal power failure
Material
Work Zone: 304 stainless steel
Main Body: Cold-rolled steel with anti-bacteria powder coating
Work Surface Height
770mm(can be customized)
800mm(Can be customized)
Caster
Footmaster caster
Power Supply
AC220V, 50/60Hz; 110V, 60Hz
Standard Accessory
LED lamp 2pcs, UV lamp 2pcs, Base stand, Remote control, Foot switch, Drain valve, Waterproof socket 2pcs, Exhaust blower, Exhaust duct
Optional Accessory
Water and gas tap, Electric height adjustable base stand, Armrest
Gross Weight
243kg; 52kg
282kg; 64kg
300kg; 64kg
408kg; 64kg
Package
Main Body Exhaust Blower(W*D*H)
1230*990*1840mm
1430*990*1840mm
1630*990*1840mm
1930*990*1840mm
930*800*700mm
930*860*740mm
930*860*740mm
930*860*740mm
1. Color key display: large screen, more clear and dynamic display of various values of the safety cabinet, and has the function of appointment timing.
12. The one-piece hand rest has a large fat surface, which has a larger contact area with the arm and is more comfortable.

Model
BSC-1100IIB2-X
BSC-1300IIB2-X
BSC-1500IIB2-X
BSC-1800IIB2-X
External Size(W*D*H)
1100*775*2250mm
1300*775*2250mm
1500*775*2250mm
1800*775*2250mm
Internal Size(W*D*H)
940*600*660mm
1150*600*660mm
1350*600*660mm
1625*600*660mm
Tested Opening
Safety height 200mm(8")
Max Opening
480mm
Inflow Velocity
0.53±0.025m/s
Downflow Velocity
0.33±0.025m/s
ULPA Filter
Two, 99.9995% efficiency at 0.12μm, filter life indicator
Front Window
Motorized. Two-layer laminated toughened glass≥6mm. Anti UV
Noise
≤65dB
UV Lamp
30W*1
30W*1
40W*1
40W*1
UV timer, UV life indicator, emission of 235.7 nanometers for most efficient decontamination
LED Lamp
12W*2
14W*2
16W*2
16W*2
Illumination
≥1000Lux
Consumption
1300W
1300W
1300W
1300W
Waterproof Socket
Two, total load of two sockets: 500W
Display
LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and downflow velocity, filter life, humidity and temperature, system working time etc.
Control System
Microprocessor
Airflow System
0% air recirculation, 100% air exhaust
Visual and Audio Alarm
Abnormal airflow velocity, Filter replacement, Front window at unsafe height, High filter pressure alarm, abnormal power failure
Material
Work Zone: 304 stainless steel
Main Body: Cold-rolled steel with anti-bacteria powder coating
Work Surface Height
770mm(can be customized)
800mm(Can be customized)
Caster
Footmaster caster
Power Supply
AC220V, 50/60Hz; 110V, 60Hz
Standard Accessory
LED lamp 2pcs, UV lamp 2pcs, Base stand, Remote control, Foot switch, Drain valve, Waterproof socket 2pcs, Exhaust blower, Exhaust duct
Optional Accessory
Water and gas tap, Electric height adjustable base stand, Armrest
Gross Weight
243kg; 52kg
282kg; 64kg
300kg; 64kg
408kg; 64kg
Package
Main Body Exhaust Blower(W*D*H)
1230*990*1840mm
1430*990*1840mm
1630*990*1840mm
1930*990*1840mm
930*800*700mm
930*860*740mm
930*860*740mm
930*860*740mm
AC Series Class II B2 Biological Safety Cabinet BSC-1100IIB2-X BSC-1300IIB2-X BSC-1500IIB2-X BSC-1800IIB2-X — professional solution in the Biological Safety Cabinet category
AC Series Class II B2 Biological Safety Cabinet BSC-1100IIB2-X BSC-1300IIB2-X BSC-1500IIB2-X BSC-1800IIB2-X is a modern business-class equipment designed for daily operation in demanding conditions. The combination of reliability, accuracy and durability makes this product the optimal choice for companies seeking consistent process quality, predictable results and reduced operating costs. The purchase price is a reasonable investment in the development and scaling of the enterprise.
Why AC Series Class II B2 Biological Safety Cabinet BSC-1100IIB2-X BSC-1300IIB2-X BSC-1500IIB2-X BSC-1800IIB2-X
- Reliability and durability: industrial-grade design and materials withstand intense loads.
- Accuracy and repeatability: stable performance that reduces the risk of defects and downtime.
- Process efficiency: optimizing staff time and increasing productivity at every stage.
- Compliance with standards: manufactured taking into account current international requirements for the business segment.
- Ease of integration: AC Series Class II B2 Biological Safety Cabinet BSC-1100IIB2-X BSC-1300IIB2-X BSC-1500IIB2-X BSC-1800IIB2-X fits easily into existing process lines and IT infrastructure.
Areas of application Biological Safety Cabinet
Biological Safety Cabinet solutions are widely used in manufacturing, research laboratories, construction sites, energy and agricultural sectors, pharmaceutical and food industries. Implementing AC Series Class II B2 Biological Safety Cabinet BSC-1100IIB2-X BSC-1300IIB2-X BSC-1500IIB2-X BSC-1800IIB2-X allows you to increase productivity, ensure accuracy of operations and optimize resource utilization across the entire enterprise.
Key benefits for business
Return on investment
Reduced service costs and downtime, predictive budget planning, fast ROI thanks to stable operation.
Scalability
Readiness to expand capacity and integrate with other solutions without re-engineering processes.
Safety and quality control
Compliance with standards, which simplifies audits and increases the trust of customers and partners.
Technical highlights
- Industrial reliability of components and assemblies for long-term trouble-free operation.
- Stable parameters that ensure repeatability of results and product quality.
- Energy efficiency and optimized life cycle cost of ownership.
- Compatibility with related equipment and standard data exchange protocols.
Service and support
Together with AC Series Class II B2 Biological Safety Cabinet BSC-1100IIB2-X BSC-1300IIB2-X BSC-1500IIB2-X BSC-1800IIB2-X you receive professional advice on selection, commissioning and personnel training. We provide warranty and post-warranty service, prompt delivery of consumables and spare parts. Our team supports projects from the procurement stage to the stable operation of the complex so that your business receives the expected result every day.
To select a package for your needs or to receive a commercial offer, add AC Series Class II B2 Biological Safety Cabinet BSC-1100IIB2-X BSC-1300IIB2-X BSC-1500IIB2-X BSC-1800IIB2-X to the cart or contact our manager. We will help integrate Biological Safety Cabinet into existing processes without downtime and unnecessary costs.
No questions about this product, be the first and ask your question.